p.5992025 Nonresidential Appendices (b) Confirm that the fan system power at 60% design airflow rate [Step 4(a)] is no
p.599greater than 40% of the exhaust fan system airflow rate at maximum wind speed [Step 1(d)].
p.599Fume Hood Automatic Sash Closure System Acceptance Test
p.599NA7.17.1 Construction Inspection
p.599Verify and document the following prior to functional testing:
p.599a) The fume hood sash zone presence sensor has a valid factory calibration
p.599certificate.
p.599b) Each fume hood sash obstruction sensor has a valid factory calibration certificate.
p.599c) Presence sensor has been located and adjusted to minimize false signals.
p.599d) Presence sensor pattern does not enter adjacent zones.
p.599e) Sash obstruction sensor has been installed per manufacturer instructions.
p.599f) Presence sensor has been installed per manufacturer instructions.
p.599NA7.17.2 Functional Testing
p.599For each sash closure control system to be tested, perform the following:
p.599a) Test auto close operation. Verify and document the following:
p.5991. Open the sash to maximum position or sash stop, whichever is lower.
p.5992. Vacate zone presence sensor range to simulate unoccupied state and confirm that sash closes automatically to minimum, closed position within 5 minutes.
p.5993. Verify that the presence sensor does not trigger a false signal from movement in an area adjacent to the space containing the controlled sash.
p.599b) Confirm that the manual controls are operational. Verify and document the
p.599following:
p.599Open Test
p.5991. If equipped, disable any auto open control mode.
p.5992. Close sash to its minimum, closed position and confirm that it does not open automatically with triggering of the zone presence sensor.
p.5993. If equipped, open the sash using a push button, foot pedal, or similar mechanism, confirming that the sash raises to the maximum position or sash stop. Otherwise, manually open the sash by hand.
p.599Closed Test
p.5991. If equipped, press the button that closes the sash and ensure that the sash closes to the minimum, closed height. Otherwise, close by hand.
p.599Appendix NA7 – Installation and Acceptance Requirements for Nonresidential Buildings and Covered Processes